| Model Number |
2450-0500 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/15/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medical device expiration date: unknown.Device manufacture date: unknown.A sample is available for evaluation.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd alaris pump module smartsite infusion set a part of the device detached from the rest without warning the following information was provided by the initial reporter: material #: 2450-0500.Batch #: unknown.The customer has experienced issues with the stopcock becoming detached from the set.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: the customer reported the stopcock detached from set on 2450-0500, and did not return a sample or picture.Despite this, a memo was provided from manufacturing based on failure description, and material.The root cause for the issue was traced to an imbalance of workload on the stopcock station.A new line layout was created on october 19, 2021 to improve the assembly sequence and prevent the imbalance on the stopcock station.A device history record review could not be performed on model 2450-0500 because the lot number is unknown.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd alaris pump module smartsite infusion set a part of the device detached from the rest without warning the following information was provided by the initial reporter: material #: 2450-0500, batch #: unknown.The customer has experienced issues with the stopcock becoming detached from the set.
|
| |
|
Search Alerts/Recalls
|