It was reported that, during an unknown procedure, a piece of the ambient hipvac wand broke off into the patient.The surgeon was able to retrieve the piece.Surgery was resumed, after a non-significant delay, with a back-up device.No patient complications were reported.
|
|
H3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.There was no way to determine if the device contributed to the reported event.The complaint was not confirmed.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.Insufficient product identification information was provided and thus a manufacturing record review could not be conducted.Based on the information available, there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.Our clinical investigation concluded: per the details of the complaint, the broke piece of the ambient hipvac was retrieved from inside of the patient by the surgeon.It was reported, the procedure was completed with a non-significant delay with a back-up device.Since there were no complications reported, no further clinical/medical assessment is warranted at this time.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|