• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAYLIS MEDICAL COMPANY INC. VERSACROSS RF WIRE; CATHETER, SEPTOSTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAYLIS MEDICAL COMPANY INC. VERSACROSS RF WIRE; CATHETER, SEPTOSTOMY Back to Search Results
Model Number VXW-35-230-P
Medical Device Problem Code Deformation Due to Compressive Stress (2889)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/27/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
While there was no patient injury reported, baylis medical company inc.Has decided to report this event due to the 30-minute procedural delay.Dhr review was completed for the lot in question, and it was confirmed that all devices met relevant requirements prior to release.
 
Event or Problem Description
A 30-minute procedural delay was reported due to kinking of the versacross rf wire (rf wire) after multiple attempts were made to cross a thick calcified septum.Following this issue, the rf wire was removed from the patient together with the versacross sheath/dilator assembly.Upon inspection of the devices, damage to the rf wire was noted.The procedure was able to continue after a new versacross rf wire was used to perform a successful transseptal puncture.While there was no patient injury reported, baylis medical company inc.Has decided to report this event due to the procedural delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSACROSS RF WIRE
Common Device Name
CATHETER, SEPTOSTOMY
Manufacturer (Section D)
BAYLIS MEDICAL COMPANY INC.
5959 trans-canada highway
montreal, quebec H4T 1 A1
CA  H4T 1A1
Manufacturer (Section G)
BAYLIS MEDICAL COMPANY INC.
2775 matheson blvd. east
mississauga, ontario L4W 4 P7
CA   L4W 4P7
Manufacturer Contact
meghal khakhar
2775 matheson blvd. east
mississauga, ontario L4W 4-P7
CA   L4W 4P7
MDR Report Key13624931
Report Number9710452-2022-00011
Device Sequence Number5060304
Product Code DXF
UDI-Device Identifier00685447003311
UDI-Public(01)00685447003311(17)240601(10)VXFF310521
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K150709
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/01/2024
Device Model NumberVXW-35-230-P
Device Lot NumberVXFF310521
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/31/2022
Initial Report FDA Received Date02/28/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/10/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-