| Model Number |
VXW-35-230-P |
| Medical Device Problem Code |
Deformation Due to Compressive Stress (2889)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/27/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
While there was no patient injury reported, baylis medical company inc.Has decided to report this event due to the 30-minute procedural delay.Dhr review was completed for the lot in question, and it was confirmed that all devices met relevant requirements prior to release.
|
| |
|
Event or Problem Description
|
|
A 30-minute procedural delay was reported due to kinking of the versacross rf wire (rf wire) after multiple attempts were made to cross a thick calcified septum.Following this issue, the rf wire was removed from the patient together with the versacross sheath/dilator assembly.Upon inspection of the devices, damage to the rf wire was noted.The procedure was able to continue after a new versacross rf wire was used to perform a successful transseptal puncture.While there was no patient injury reported, baylis medical company inc.Has decided to report this event due to the procedural delay.
|
| |
|