• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS IOL; LENS, MULTIFOCAL INTRAOCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS IOL; LENS, MULTIFOCAL INTRAOCULAR Back to Search Results
Model Number DFW150
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Visual Disturbances (2140); Halo (2227); Discomfort (2330)
Date of Event 11/30/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the after the intraocular lens (iol) was implanted in the left eye, the patient complained of nighttime halos, glare, seeing individual rings, and colors are off and not sharp.The patient cannot read for any extended period of time, however near vision is fine for short periods.Intermediate and distance vision are worse than prior to surgery.The symptoms significantly interfered with the patient's daily activities and were debilitating.The iol was explanted.Another johnson and johnson iol (different model, diu150, same diopter) was implanted as replacement.There was no medical or surgical intervention required and no patient injury.The patient was happy at one day post-operative of the iol exchange.The patient's pre-operative best corrected and uncorrected visual acuity on (b)(6) 2021 was 20/25.The patient's post-operative uncorrected visual acuity was 20/30 and post-operative best corrected visual acuity was 20/20 on (b)(6) 2022.There are no additional interventions planned.No further information was provided.
 
Additional Manufacturer Narrative
The device was not returned for analysis.Therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Additional information: section d9: device available for evaluation? yes.Section d9: returned to manufacturer: yes.Section d9: date returned to manufacturer: feb 28, 2022.Section h3: evaluated by manufacturer: yes.Device evaluation: visual inspection under magnification revealed that the lens was received cut in half, which is consistent with a lens that was handled during explant.No issues related to or could have contributed to visual issues were observed.No handpiece was received as part of this return therefore no product evaluation could be performed on the handpiece.Based on the return condition of the lens, the complaint issue could not be confirmed, and no product deficiency could be identified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
LENS, MULTIFOCAL INTRAOCULAR
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key13625329
Report Number3012236936-2022-00532
Device Sequence Number17402628
Product Code MFK
UDI-Device Identifier05050474751248
UDI-Public(01)05050474751248(17)241009
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/09/2024
Device Model NumberDFW150
Device Catalogue NumberDFW150U190
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/02/2022
Supplement Date Received by Manufacturer04/05/2022
Initial Report FDA Received Date02/28/2022
Supplement Report FDA Received Date04/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
Patient Weight69 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-