Model Number M00513401
Medical Device Problem Code Device Contaminated During Manufacture or Shipping (2969)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/31/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that a radial jaw 4 standard capacity biopsy forceps was to be used in a biopsy procedure performed on (b)(6) 2022.During preparation, the physician noticed that there was something like wood shavings inside the sterile area of the packaging.Another radial jaw 4 biopsy forceps was used to complete the procedure.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Initial reporter phone: (b)(6).Device code: (b)(4).The device has been received for analysis; however, the analysis has not yet been completed.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Additional Manufacturer Narrative
Block e1: initial reporter phone: (b)(6).Initial reporter fax: (b)(6).Block h6: device code a1802 captures the reportable event of packaging foreign matter.Block h10: the returned radial jaw 4 sc biopsy forceps was analyzed, and a visual evaluation noted that the pouch was completely unopen.A foreign material was found inside the pouch.A dimensional test confirmed that the foreign matter inside is greater than the tolerated size.No other issues with the device were noted.The reported event was confirmed.Based on all available information, the problem was traced to manufacturing process.It is concluded that the foreign material is inside the pouch due to a deficiency in the process.Therefore, the root cause chosen is manufacturing deficiency.An investigation to address this issue is in progress.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
 
Event or Problem Description
It was reported to boston scientific corporation that a radial jaw 4 standard capacity biopsy forceps was to be used in a biopsy procedure performed on (b)(6) 2022.During preparation, the physician noticed that there was something like wood shavings inside the sterile area of the packaging.Another radial jaw 4 biopsy forceps was used to complete the procedure.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RADIAL JAW 4
Common Device Name
FORCEPS, BIOPSY, NON-ELECTRIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
pmt 741 persiaran cassia selat
bandarcassia, pulau pinan 14110
MY   14110
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key13625390
Report Number3005099803-2022-00956
Device Sequence Number17424540
Product Code FCL
UDI-Device Identifier08714729777960
UDI-Public08714729777960
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/20/2022
Device Model NumberM00513401
Device Catalogue Number08714729777960
Device Lot Number0023988610
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/22/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/02/2022
Supplement Date Received by Manufacturer03/16/2022
Initial Report FDA Received Date02/28/2022
Supplement Report FDA Received Date04/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/21/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-