Additional information - this emdr is being submitted to include the below: d4: unique identifier (udi) # h6: investigation results under imdrf cause investigation code, imdrf investigation findings, imdrf cause conclusions.Batch record review: the batch 1l01125 was manufactured on 11/11/2021, form fill and seal #1 manufacturing line, complaint investigator performed a batch record review on 30/dec/2022 to verify if all the applicable procedures were followed and no issues were found; all the components for assembly were correct per bill of materials (bom) and all the tooling information documented was also correct, system application product (sap) material 1709735 and manufacturing order (b)(4).The batch record review supports that there were no discrepancies related to the issue reported.Photograph, video and/or physical sample evaluation: no photographs associated with this case was received.No return sample has been received for this complaint.Complaint data analysis: on 30/dec/2022, complaint investigator performed a search of complaints into database for lot number 1l01125, and as a result, a total of 01 complaint was found and only 01 is related to the same malfunction code of this investigation therefore, no potential trend is observed.As per complaint manufacturing investigation procedure work instruction (wi), it is not required to open a nonconformance report (ncr) for type 2 complaints which were not confirmed and no potential trend is identified.Per product harm and reportability index (phri) and medical device reporting (mdr) evaluation there was no product malfunction reported with this complaint.No additional action is required, and this complaint will be closed.This issue will be monitored through the post market product monitoring review process, standard operating procedure (sop).To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number reporting site: 1049092 manufacturing site: 9618003.
|