INTUITIVE SURGICAL, INC DA VINCI; ENDOSCOPE Back to Search Results
Model Number 372010-02
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Intuitive surgical, inc.(isi) received the 0 degree endoscope and completed the device evaluation.The endoscope was received with the mechanical damage to distal window assembly hermetic seal resulting in fluid invasion and subsequent major damage to the optical components.Fragments of the adhesive of the distal window are missing.A review of the site's complaint history found that there were no other complaints for this product or this event.No image or video of the procedure was provided for review.Verification of the event details cannot be performed via system logs because system logs do not exist for da vinci si endoscopes.This complaint is being reported based on the following: failure analysis acknowledges that a fragment was missing from a portion of the device that enters the patient.Based on the information provided, there was no report from the customer that a fragment from the instrument fell into the patient during the procedure.While there was no report of patient involvement, the reported failure mode could likely cause or contribute to an adverse event if the fragment(s) did fall into the patient.
 
Event or Problem Description
The endoscope was returned with no reported complaint.There was no report of patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DA VINCI
Common Device Name
ENDOSCOPE
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key13625820
Report Number2955842-2022-10406
Device Sequence Number12071292
Product Code NAY
UDI-Device Identifier00886874110447
UDI-Public(01)00886874110447
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
K101077
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 02/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number372010-02
Device Catalogue Number372010
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/28/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/02/2022
Initial Report FDA Received Date02/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DA VINCI INSTRUMENTS AND ACCESSORIES
Patient SexUnknown
-
-