MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® GOODWIN SOUND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® GOODWIN SOUND Back to Search Results
Model Number 042812
Medical Device Problem Code Corroded (1131)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/15/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported event was confirmed however the cause was unknown.One sample was confirmed to exhibit the reported failure.The device had not met specifications.As the device was not used on a patient it was not used for diagnostic or treatment purposes.A potential root cause for this failure could be "improper high level disinfection or sterilization".A potential root cause for this failure could be "improper high level disinfection or sterilization".The lot number was unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "visual inspection examine each device for cleanliness.If visible soil remains, repeat manual cleaning instructions.Verify that devices are free of deformations (e.G.Bent, broken, corroded, cracked, worn, or fractured) that may impact its intended use.Do not use the product if it is damaged or defective.".
 
Event or Problem Description
It was reported that the customer received a partial order of quite a few goodwin sound replacements as it had consistently had issues with rust 042822 and 042820 and discoloration 042816and 042814 in the lumen.It was also stated upon receiving the new replacements, they were inspected with a borescope and discoloration was noted prior to any reprocessing.Per follow up via email on 16aug21, the lot numbers of the affected sounds which are mixed with new box and never processed.The new box, which was processed initially, never used on patient.This was the ongoing issue for the customer and consistently replacing these.The customer checked and verified internal process and follow the ifu verbatim, flushing with sterile water at point of use.The box and packaging that customer received was for 11 sounds.They have total of 24 sounds 042824 and 042826.
 
Additional Manufacturer Narrative
The reported event was confirmed however the cause was unknown.One sample was confirmed to exhibit the reported failure.The device had not met specifications.As the device was not used on a patient it was not used for diagnostic or treatment purposes.A potential root cause for this failure could be "improper high level disinfection or sterilization".A potential root cause for this failure could be "improper high level disinfection or sterilization".The lot number was unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "visual inspection examine each device for cleanliness.If visible soil remains, repeat manual cleaning instructions.Verify that devices are free of deformations (e.G.Bent, broken, corroded, cracked, worn, or fractured) that may impact its intended use.Do not use the product if it is damaged or defective.".
 
Event or Problem Description
It was reported that the customer received a partial order of quite a few goodwin sound replacements as it had consistently had issues with rust 042822 and 042820 and discoloration 042816and 042814 in the lumen.It was also stated upon receiving the new replacements, they were inspected with a borescope and discoloration was noted prior to any reprocessing.Per follow up via email on 16aug21, the lot numbers of the affected sounds which are mixed with new box and never processed.The new box, which was processed initially, never used on patient.This was the ongoing issue for the customer and consistently replacing these.The customer checked and verified internal process and follow the ifu verbatim, flushing with sterile water at point of use.The box and packaging that customer received was for 11 sounds.They have total of 24 sounds 042824 and 042826.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® GOODWIN SOUND
Common Device Name
GOODWIN SOUND
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key13625907
Report Number1018233-2022-00943
Device Sequence Number20130346
Product Code FBX
UDI-Device Identifier00801741047220
UDI-Public(01)00801741047220
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number042812
Device Catalogue Number042816
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/07/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/03/2022
Initial Report FDA Received Date02/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-