Model Number N560
Medical Device Problem Code Noise, Audible (3273)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during service and repair, the unit speaker was defective.It was stated that when the unit was turned on there was no sound and found that one of the cables were broken.The speaker was replaced with a new one.
 
Additional Manufacturer Narrative
Medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the speaker in the monitor was defective.The issue was resolved by replacing a new speaker.It was reported that the speaker in the monitor was defective.The reported issue was confirmed.The most likely cause was traced to a component failure.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE OXIMETER OXIMAX
Common Device Name
GASTROINTESTINAL TUBES WITH ENTERAL SPECIFIC C
Manufacturer (Section D)
MEDIANA CO. LTD.
1650 1 2 donghwa ri munmak eu
kangwon do wonju si 00000
KS  00000
Manufacturer (Section G)
MEDIANA CO. LTD.
1650 1 2 donghwa ri munmak eu
kangwon do wonju si 00000
KS   00000
Manufacturer Contact
clay chandler
5920 longbow drive
boulder, CO 80301
3035306409
MDR Report Key13626809
Report Number2936999-2022-00161
Device Sequence Number20146447
Product Code PIF
UDI-Device Identifier08056093720976
UDI-Public08056093720976
Combination Product (Y/N)N
Initial Reporter CountryTU
PMA/510(K) Number
K181787
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN560
Device Catalogue NumberN560
Device Lot NumberCX07113284
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/03/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/09/2022
Supplement Date Received by Manufacturer04/27/2022
Initial Report FDA Received Date03/01/2022
Supplement Report FDA Received Date05/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-