As the lot number for the device was provided, a review of the device history record is currently being performed.The return of the sample is pending.However, photos were provided for review.The investigation of the reported event is currently underway.(expiry date: 10/2024) the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.Device pending return.
|
|
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical device was not returned for evaluation.Three photos were provided and reviewed.The first photo shows the catheter which is noted to have a circumferential break proximal to the balloon exposing the inner guide wire lumen.The second photo shows the catheter which is noted to have a circumferential break proximal to the balloon.The third photo shows the catheter which is noted to have a circumferential break proximal to the balloon exposing the inner guide wire lumen.No other specific anomalies noted.Therefore, the investigation is confirmed for the reported catheter break as a circumferential break can be seen proximal to the balloon from the provided photos.A definitive root cause for the alleged break could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 10/2024), g3.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|