| Model Number |
24700 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/17/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 85% stenosed target lesion was located in the moderately tortuous and moderately calcified external iliac artery.A 8.0mmx60mmx135cm sterling balloon catheter was advanced for dilation.However, on the first inflation at 10 atmospheres for 10 seconds, the balloon ruptured.The device was removed and the procedure was completed with another of the same device.There were no patient complications nor injuries reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: returned product consisted of a sterling balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed no damages.Microscopic examination revealed a pinhole 2.4cm from the tip.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 85% stenosed target lesion was located in the moderately tortuous and moderately calcified external iliac artery.A 8.0mmx60mmx135cm sterling balloon catheter was advanced for dilation.However, on the first inflation at 10 atmospheres for 10 seconds, the balloon ruptured.The device was removed and the procedure was completed with another of the same device.There were no patient complications nor injuries reported.
|
| |
|