H10: manufacturing review: a manufacturing review was not requested as the lot number reported is unknown.Investigation summary: one 12fr hickman d/l catheter was returned for evaluation.Gross visual, microscopic, tactile and functional testing were performed.The investigation is confirmed for the reported leak, fracture, obstruction of flow and identified burst issues as a c-shaped split was noted on the white luer extension leg approximately 1.0cm from the clamping sleeve.A longitudinal split was noted on the red luer extension leg approximately 0.6cm from the clamping sleeve.During tactile evaluation, extension legs and catheter were gently stretched and necking was observed throughout.All areas were noted to be normally elastic.Upon infusion of the white luer, a leak from the c-shaped split was noted while water exited the distal end of the catheter; only air was aspirated into the syringe.Thrombus was observed exiting the catheter on both infusion attempts.The definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: b5, g3, h6(device, method).H11: h6(result, conclusion).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|