• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number RLT281416
Medical Device Problem Codes Patient-Device Incompatibility (2682); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Aneurysm (1708)
Date of Event 06/19/2018
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient presented on (b)(6) 2016 with an abdominal aortic aneurysm (pre-implant ct diameter 54 mm) that was treated with the implantation of gore® excluder® aaa endoprostheses.After an initial decrease of the aneurysm diameter (ultrasound 49 mm) in 2017, minor aneurysm growth was visible on (b)(6) 2018 (ultrasound 52 mm).On (b)(6) 2021, a type ii endoleak originating from lumbar arteries was identified that contributed to additional significant aneurysm growth (ultrasound diameter 66 mm).On (b)(6) 2021, the endoleak was successfully treated with embolization of the lumbar arteries.The patient tolerated the procedure.
 
Additional Manufacturer Narrative
Device remains implanted in patient and is unavailable for evaluation.Per gore® excluder® aaa endoprosthesis instructions for use (ifu), adverse events that may occur and/or require intervention include, but are not limited to: endoleak.A review of the manufacturing records for the device verified that the lot met all prerelease specifications.
 
Additional Manufacturer Narrative
H3: device remains implanted in patient and is unavailable for evaluation.H6: according to the gore® excluder® aaa endoprosthesis instructions for use (ifu), users are made aware of the risks associated with type ii endoleaks and are instructed to consider the risks and benefits discussed in the ifu for each patient before using the devices.Additionally, per ifu, adverse events that may occur and / or require intervention include, but are not limited to endoleak and aneurysm enlargement.A review of the manufacturing records for the device verified that the lot met all prerelease specifications.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® AAA ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL PHOENIX 2 B/P
32470 n. north valley parkway
phoenix AZ 85085
Manufacturer Contact
tom hormby
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key13630591
Report Number3007284313-2022-01797
Device Sequence Number17412154
Product Code MIH
UDI-Device Identifier00733132622542
UDI-Public00733132622542
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Study
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/04/2019
Device Model NumberRLT281416
Device Catalogue NumberRLT281416
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/31/2022
Supplement Date Received by Manufacturer01/31/2022
Initial Report FDA Received Date03/01/2022
Supplement Report FDA Received Date08/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/04/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PXC141200 / (B)(6).
Outcome Attributed to Adverse Event Other;
Patient SexMale
Patient Weight61 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-