• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC TRILOGY 200; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC TRILOGY 200; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 1032804
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/17/2015
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This mdr is being submitted as part of a batch submission of previously closed complaints that were reassessed as reportable foam degradation complaints; discovered as part of a retrospective remediation review.
 
Event or Problem Description
A ventilator was returned to the manufacturer for service.There was no harm or injury reported.During the evaluation of the device at the manufacturer's service center, the air inlet path assembly was replaced due to sponge peeled off and pollen filter and air path foam were replaced.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Additional Manufacturer Narrative
The manufacturer previously reported the investigation was ongoing and a follow up report would be submitted when the investigation was complete.No follow up report is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY 200
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
head of quality src
6501 living place
pittsburgh, PA 15206
2763970028
MDR Report Key13634055
Report Number2518422-2022-06110
Device Sequence Number5039710
Product Code CBK
UDI-Device Identifier00606959033863
UDI-Public00606959033863
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K093416
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 04/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1032804
Device Catalogue Number1032804
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/20/2015
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/17/2015
Supplement Date Received by Manufacturer06/17/2015
Initial Report FDA Received Date03/01/2022
Supplement Report FDA Received Date04/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/16/2010
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
FDA RES 88071
Patient Sequence Number1
Patient SexUnknown
-
-