| Model Number |
23295 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 90% stenosed target lesion was located near a shunt anastomosis in a moderately tortuous and severely calcified vessel.A 3.5-4/4t/90 symmetry balloon catheter was advanced for dilatation.However, during the initial inflation at 12 atmospheres for 30 seconds, the balloon ruptured and a vertical tear was noted.The device was removed and the procedure was completed with a different device.No patient complications were reported and the patient was in good condition post-procedure.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 90% stenosed target lesion was located near a shunt anastomosis in a moderately tortuous and severely calcified vessel.A 3.5-4/4t/90 symmetry balloon catheter was advanced for dilatation.However, during the initial inflation at 12 atmospheres for 30 seconds, the balloon ruptured and a vertical tear was noted.The device was removed and the procedure was completed with a different device.No patient complications were reported and the patient was in good condition post-procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
E1: initial reporter address 1: (b)(6).Device evaluated by mfr.: a symmetry device was returned for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.A longitudinal tear was identified in the balloon material.The tear measured 7mm in length and extended from a position 3mm distal of the proximal markerband to 3mm proximal of the proximal markerband.A visual and tactile examination identified no damage or issues with the shaft of the device.A visual examination identified no issues with the tip of the device that could have contributed to the complaint incident.This concludes the product analysis.
|
| |
|
Search Alerts/Recalls
|