Model Number MN60AC
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/11/2021
Type of Reportable Event Serious Injury
Event or Problem Description
A physician reported during the intraocular lens implantation the haptic was broken.The lens was implanted and explanted on the same day.There was patient contact but no harm to the patient.The surgery was completed on the next day.Additional information has been requested.
 
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.There have been no other complaints reported in the lot number.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
Additional information provided in d.9.H.3., h.6., and h.10.The lens was returned for evaluation.Solution is dried on the lens.One haptic is broken and bent in the distal area (not returned).The file indicates the use of a qualified cartridge and handpiece.The file also indicates the use of a non-qualified viscoelastic.Broken haptic damage was observed.All lenses are 100% inspected for cosmetic attributes and the damage exhibited by the returned complaint sample would not have met company release criteria.Based on our observation, it can be reasonably concluded that the damage is not manufacturing related.Due to the presence of surgical solution and the condition of the returned sample, the damage is most likely related to customer handling.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF MULTIPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key13634993
Report Number1119421-2022-00370
Device Sequence Number2213502
Product Code HQL
UDI-Device Identifier00380655095218
UDI-Public00380655095218
Combination Product (Y/N)N
Initial Reporter CountryMY
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/30/2021
Device Model NumberMN60AC
Device Catalogue NumberMN60AC.250
Device Lot Number12512901
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/07/2022
Supplement Date Received by Manufacturer03/31/2022
Initial Report FDA Received Date03/01/2022
Supplement Report FDA Received Date04/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/12/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DISCOVISC; MONARCH II CARTRIDGE B; MONARCH III INJECTOR
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-