• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEUWAVE MEDICAL, INC. NEUWAVE PR PROBE 15CM 17GA; SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEUWAVE MEDICAL, INC. NEUWAVE PR PROBE 15CM 17GA; SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES Back to Search Results
Model Number PR15
Medical Device Problem Code Insufficient Heating (1287)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/20/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch #: ml20065362 investigation summary: there is no call home available.The probe returned with the cannula broken off from the handle.The root cause is unknown since no further information into how the break happened was included in the additional information.Lot ml20065362 was reviewed (in process sn 7).Manufacture date: 7/27/20.Expiration date: 3/1/24.There were no problems seen during the manufacturing and testing of this lot.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that at the end of the ablation, the temperature dropped to 0 degrees, and it was not clear whether the ablation was completely achieved with the probe.A new probe was opened to complete the ablation with the desired result.No harm to patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUWAVE PR PROBE 15CM 17GA
Common Device Name
SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES
Manufacturer (Section D)
NEUWAVE MEDICAL, INC.
3529 anderson st
madison 53704
Manufacturer (Section G)
NEUWAVE MEDICAL, INC.
3529 anderson st
madison 53704
Manufacturer Contact
kara ditty-bovard
3529 anderson st
madison, WI 53704
6107428552
MDR Report Key13635781
Report Number3008769756-2022-00008
Device Sequence Number2124605
Product Code NEY
UDI-Device Identifier00853719006395
UDI-Public00853719006395
Combination Product (Y/N)N
PMA/510(K) Number
K160936
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/01/2024
Device Model NumberPR15
Device Catalogue NumberPR15
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/26/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/10/2022
Initial Report FDA Received Date03/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/27/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-