| Model Number |
24704 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/09/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The target 70% stenosed target lesion was located in the severely tortuous and severely calcified dorsalis pedis artery.A 1.5mm x 40mm x 150cm, otw coyote balloon catheter was advanced for dilation.However, during second inflation at 14 atmospheres for 10 minutes, the balloon burst.The procedure was cancelled/rescheduled.There were no patient complications nor injuries reported and the patient condition was good.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The target 70% stenosed target lesion was located in the severely tortuous and severely calcified dorsalis pedis artery.A 1.5mm x 40mm x 150cm, otw coyote balloon catheter was advanced for dilation.However, during second inflation at 14 atmospheres for 10 minutes, the balloon burst.The procedure was cancelled/rescheduled.There were no patient complications nor injuries reported and the patient condition was good.
|
| |
|
Additional Manufacturer Narrative
|
|
Returned product consisted of a coyote es balloon catheter.The outer shaft, inner shaft, balloon, and tip were visually and microscopically examined.The batch number was confirmed correct.Microscopic analysis confirmed a pinhole 4 cm from the distal end of the balloon.Device analysis was able to confirm the reported complaint of a material rupture.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The target 70% stenosed target lesion was located in the severely tortuous and severely calcified dorsalis pedis artery.A 1.5mm x 40mm x 150cm, otw coyote balloon catheter was advanced for dilation.However, during second inflation at 14 atmospheres for 10 minutes, the balloon burst.The procedure was cancelled/rescheduled.There were no patient complications nor injuries reported and the patient condition was good.
|
| |
|