• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON LABORATORIES IRELAND LTD. ACRYSOF SINGLEPIECE IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON LABORATORIES IRELAND LTD. ACRYSOF SINGLEPIECE IOL; INTRAOCULAR LENS Back to Search Results
Model Number SA60AT
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Endophthalmitis (1835)
Date of Event 02/05/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported that a patient experienced endophthalmitis in the operated eye that was noted during the post-operative check-up.Additional information has been requested but not yet received.
 
Additional Manufacturer Narrative
The product was not returned for analysis.The root cause for the reported complaint could not be determined.Based on the results from the product history record, the products met release criteria.There have been no other complaints for this lot.Investigation has been completed based on current information.No further action is warranted at this time.Based on our trend analysis, there are no adverse trends identified for this complaint and based on these findings the residual risk remains at an acceptable level.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF SINGLEPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key13635995
Report Number9612169-2022-00094
Device Sequence Number5039773
Product Code HQL
UDI-Device Identifier00380652555807
UDI-Public00380652555807
Combination Product (Y/N)N
Initial Reporter CountryPL
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/14/2026
Device Model NumberSA60AT
Device Catalogue NumberSA60AT.200
Device Lot Number25250383
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/07/2022
Supplement Date Received by Manufacturer05/16/2022
Initial Report FDA Received Date03/01/2022
Supplement Report FDA Received Date05/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CENTURION SURGICAL PROCEDURE PAK; CENTURION VISION SYSTEM; PROVISC OPHTHALMIC VISCOSURGICAL DEVICE; UNSPECIFIED BALANCED SALT SOLUTION (BSS)
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-