• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 ALIGN® TO URETHRAL SUPPORT SYSTEM; ALIGN® TO URETHRAL SUPPORT SYSTEM - HALO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 ALIGN® TO URETHRAL SUPPORT SYSTEM; ALIGN® TO URETHRAL SUPPORT SYSTEM - HALO Back to Search Results
Model Number BRD500HL
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Apnea (1720); Asthma (1726); Erosion (1750); Fatigue (1849); Hemorrhage/Blood Loss/Bleeding (1888); High Blood Pressure/ Hypertension (1908); Low Blood Pressure/ Hypotension (1914); Incontinence (1928); Unspecified Infection (1930); Nausea (1970); Pain (1994); Urinary Retention (2119); Urinary Tract Infection (2120); Hernia (2240); Distress (2329); Prolapse (2475); Hematuria (2558); Cancer (3262); Cramp(s) /Muscle Spasm(s) (4521)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reported event could not be confirmed.No sample was returned for evaluation and the lot number is unknown.Based on the investigation findings, current manufacturing controls are considered adequate as to detect and segregate any non-conforming unit.Event described could not be confirmed as a manufacturing related issue.The alleged event is most likely associated with possible procedural/surgical complications.The lot number is unknown; therefore, the device history record could not be reviewed.The ifu currently addresses potential risks associated with surgically implantable materials which states the following: precautions the usual precautions associated with urological procedures should be followed: based on physician experience and training, a thorough assessment of each patient should be made to determine their suitability for the implant procedure.Additional consideration should be given to use of the implant in patients with a compromised immune system, any condition that would compromise healing, or any patient with a history of prior abdominal or pelvic surgeries.Consideration should also be given to the ability of the patient to tolerate the surgical procedure.Accepted surgical practice and precautions must be followed for the management of contaminated or infected wound sites, when the align® urethral support system is used.Postoperative bleeding may occur in some patients and must be controlled prior to patient release.The implant procedure requires diligent attention to anatomical structures and care to avoid puncture of large vessels, nerves, bladder, bowel, urethra and any viscera, during introducer passage.Cystoscopy should be performed to confirm bladder integrity or recognize a bladder perforation.Due to anatomical distortion that can be caused by pelvic organ prolapse, if the patient requires cystocele repair, it should be performed prior to the implantation of the sub-urethral sling.Proper placement of the sling implant at mid-urethra requires that it lie flat with minimal or no tension under the urethra.The align® urethral support system is intended as a single-use device.Do not re-sterilize any portion of the align® urethral support system.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/ or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Patients should be advised that pregnancy following a mesh sling implant procedure may negatively affect the success of the previous implant procedure and incontinence may recur.The safety and effectiveness of the align® urethral support system implant procedure has not been established for the treatment of stress urinary incontinence in males or children under the age of 18.Check the integrity of the packaging before use.Do not use the align® urethral support system if the packaging is opened or damaged.As for any implantable material, it is recommended to open the package at the time of implantation.Post-operatively the patient should be advised to refrain from heavy lifting, exercise (e.G.Cycling, jogging) and/or intercourse until the physician determines it is suitable for the patient to return to normal activities.Adverse events complications associated with the proper implantation of the align® urethral support system may include, but are not limited to: - postoperative hematoma, seroma, abscess or fistula formation, or scarring which may occur following the implant procedure.Urinary retention, bladder outlet obstruction and other voiding dysfunctions.These conditions may be associated with overcorrection/ too much tension placed on the implant.Perforations or lacerations of vessels, nerves, bladder, bowel, urethra, or any viscera, which may occur during the implantation procedure.Irritation at the operative wound site which may elicit a foreign body response that leads to wound dehiscence, inflammation and/or infection.Extrusion through vaginal epithelium or erosion into surrounding viscera and/or mucosa.Inflammation, sensitization, pain, dyspareunia, scarification, contraction, device migration and failure of the procedure resulting in recurrence of incontinence.2348, 2120, 1750, 1994 ="l", 2558, 2371, 2329 ="nl".
 
Event or Problem Description
The patient's attorney alleged a deficiency against the device.Additional information has been requested, but not yet received.Per additional information received via claim record on (b)(6) 2022, the patient has experienced a recurrent urinary tract infection, hematuria, bleeding, erosion of the vaginal mesh, mental and physical pain, disability, suffering, permanent injuries, substantial physical deformity, severe emotional distress and required additional surgical and non-surgical interventions.
 
Event or Problem Description
Per additional information received via medical records on 26apr2022, the patient has experienced prolapsed bladder, a tear in her vaginal wall, pelvic pain, vaginal pain, severe low back pain, rectocele, abnormal bladder control, significant exposure/erosion, hematuria, recurrent stress urinary incontinence, infection, pelvic pressure, cramping, clotting, cancer, connective tissue disorder, hernias, hypotension or low blood pressure, immune system disease or dysfunction including hiv/aids, pelvic trauma, pelvic tumors or fibroids, urinary retention, wound healing problems and required additional surgical and non-surgical interventions.
 
Additional Manufacturer Narrative
1928, 1930, 2119, ="l".2475, 2193, 1760, 2240, 1914="nl".H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
Per additional information received via medical records on 23may2022, the patient has experienced kidney function, hematuria, rectocele, recurrent stress urinary incontinence, recurrent urinary tract infection, vaginal bleeding, mesh exposure with granulation tissue and bleeding, prolapsed bladder, a tear in her vaginal wall, pelvic pain, vaginal pain, severe low back pain, abnormal bladder control, pelvic pressure, cramping, clotting, cancer, connective tissue disorder, hernias, hypotension or low blood pressure, immune system disease or dysfunction including hiv/aids, pelvic trauma, pelvic tumors or fibroids, urinary retention, wound healing problems, right knee pain, chronic lymphocytic leukemia, gastroesophageal reflux disease, hypertension, asthma, cervicalgia, sleep apnea, restless leg syndrome, fatigue, nausea with vomiting, raised antinuclear antibody, urgency of urination, migraine, high sodium level and required additional surgical and non-surgical interventions.
 
Additional Manufacturer Narrative
1720, 1849, 1970, 1726, 1908=nl.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALIGN® TO URETHRAL SUPPORT SYSTEM
Common Device Name
ALIGN® TO URETHRAL SUPPORT SYSTEM - HALO
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key13636436
Report Number1018233-2022-90006
Device Sequence Number5201345
Product Code OTN
UDI-Device Identifier00801741016226
UDI-Public(01)00801741016226
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K093747
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2019
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberBRD500HL
Device Catalogue NumberBRD500HL
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Event Location Hospital
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/14/2022
Supplement Date Received by Manufacturer04/26/2022
05/23/2022
Initial Report FDA Received Date03/01/2022
Supplement Report FDA Received Date05/12/2022
06/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AVAULTA SOLO ANTERIOR SYNTHETIC SUPPORT SYSTEM.; AVAULTA SOLO ANTERIOR SYNTHETIC SUPPORT SYSTEM.; AVAULTA SOLO ANTERIOR SYNTHETIC SUPPORT SYSTEM.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age54 YR
Patient SexFemale
Patient Weight77 KG
-
-