| Model Number |
00884450030813 |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
02/18/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The account alleges that during an esophogeal stent replacement due to a developed fistula of an intrinsic tumor, the original stent broke as it was being removed.The procedure was stopped as there was a need to get consent from the patient or family for an esophogeal stent and the procedure is to be completed at a later date.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device was returned for evaluation.The complaint was confirmed and the root cause is attributed to clinical use and excessive force during removal of the device.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.
|
| |
|