| Model Number |
0570-0374 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The glidescope bflex 5.0 single-use bronchoscope used in the event was returned to verathon for evaluation.On visual inspection by the verathon technical service representative they were able to confirm the reported damage; the tip of the bronchoscope was loose inside the bag it was received in.Upon completion of the evaluation the reported glidescope bflex 5.0 single-use bronchoscope was scrapped.At this time, the cause of the reported issue could not be fully determined; however, it is likely that inserting the glidescope bflex 5.0 single-use bronchoscope into a incompatible double-lumen tube may have caused or contributed to the reported issue.The "glidescope bflex bronchoscope endotracheal tube compatibility" table found on pg.16 in the glidescope bflex single-use bronchoscopes with glidescope core operations & maintenance manual (omm), currently list the use of a glidescope bflex 5.0 single-use bronchoscope with a double-lumen tube as "not supported".Any usage of the glidescope bflex 5.0 single-use bronchoscope with a double lumen-tube is considered outside of verathon's released labeling.Corrective action is not required at this time.Verathon will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
The customer reported that during a patient procedure, using a glidescope bflex 5.0 single-use bronchoscope with a double-lumen tube, a piece of plastic from the bronchoscope broke off and retained in the patient's airway.A therapeutic bronchoscopy was performed in combination with the use of bronchoscope alligator forceps / graspers to remove the broken bronchoscope piece successfully.No harm to the patient or user was reported.
|
| |
|
Search Alerts/Recalls
|