Model Number 0570-0374
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The glidescope bflex 5.0 single-use bronchoscope used in the event was returned to verathon for evaluation.On visual inspection by the verathon technical service representative they were able to confirm the reported damage; the tip of the bronchoscope was loose inside the bag it was received in.Upon completion of the evaluation the reported glidescope bflex 5.0 single-use bronchoscope was scrapped.At this time, the cause of the reported issue could not be fully determined; however, it is likely that inserting the glidescope bflex 5.0 single-use bronchoscope into a incompatible double-lumen tube may have caused or contributed to the reported issue.The "glidescope bflex bronchoscope endotracheal tube compatibility" table found on pg.16 in the glidescope bflex single-use bronchoscopes with glidescope core operations & maintenance manual (omm), currently list the use of a glidescope bflex 5.0 single-use bronchoscope with a double-lumen tube as "not supported".Any usage of the glidescope bflex 5.0 single-use bronchoscope with a double lumen-tube is considered outside of verathon's released labeling.Corrective action is not required at this time.Verathon will continue to monitor for trends.
 
Event or Problem Description
The customer reported that during a patient procedure, using a glidescope bflex 5.0 single-use bronchoscope with a double-lumen tube, a piece of plastic from the bronchoscope broke off and retained in the patient's airway.A therapeutic bronchoscopy was performed in combination with the use of bronchoscope alligator forceps / graspers to remove the broken bronchoscope piece successfully.No harm to the patient or user was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDESCOPE BFLEX 5.0 SINGLE-USE BRONCHOSCOPE
Common Device Name
BRONCHOSCOPE (FLEXIBLE OR RIGID)
Manufacturer (Section D)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA  V5C 5A9
Manufacturer (Section G)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA   V5C 5A9
Manufacturer Contact
corey kasbohm
20001 n creek pkwy
bothell, WA 98011-8218
4256295760
MDR Report Key13636467
Report Number9615393-2022-00030
Device Sequence Number20102731
Product Code EOQ
UDI-Device Identifier00879123006332
UDI-Public(01)00879123006332(10)FS211100(17)220312
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K183256
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 02/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/12/2022
Device Model Number0570-0374
Device Catalogue Number0570-0375
Device Lot NumberFS211100
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/15/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/02/2022
Initial Report FDA Received Date03/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/10/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-