Model Number 00884450386521
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Stenosis (2263)
Date of Event 12/20/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The suspect device was not returned for evaluation.The complaint could not be confirmed.The root cause could not be determined.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.
 
Event or Problem Description
The account alleges that a graft was successfully implanted in (b)(6) 2021.On (b)(6) 2022, during follow up, it was discovered that the venous outflow component had migrated out of the right atrium and is causing stenosis.The entire graft needed to be explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HERO®
Common Device Name
PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS INC.
1600 merit parkway
south jordan UT 84095
Manufacturer (Section G)
MERIT MEDICAL SYSTEMS INC.
1600 merit parkway
south jordan UT 84095
Manufacturer Contact
bryson heaton
1600 merit parkway
south jordan, UT 84095
8012084662
MDR Report Key13636548
Report Number1721504-2022-00023
Device Sequence Number5039794
Product Code DSY
UDI-Device Identifier00884450386521
UDI-Public00884450386521
Combination Product (Y/N)N
Initial Reporter CountryTC
PMA/510(K) Number
K172637
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/12/2023
Device Model Number00884450386521
Device Catalogue NumberHERO1001VOC/A
Device Lot NumberH2257524
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/24/2022
Initial Report FDA Received Date03/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-