| Model Number |
FX-410615 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Codes |
Capsular Contracture (1761); Seroma (2069)
|
| Date of Event |
08/02/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Information contained in this report was previously submitted through 410psr on 01/22/2019 and 04/22/2019.A review of the device history record has been completed.No deviations or non-conformances noted.The events of capsular contracture and seroma-late are physiological complications and analysis of the device generally does not assist allergan in determining a probable cause for these events.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.The reason for reoperation: capsular contracture baker grade iii, seroma-late.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a right-side capsular contracture, baker grade iii.Healthcare professional additionally reported right side "late seroma".Device has been explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a right-side capsular contracture, baker grade iii.Healthcare professional additionally reported right side "late seroma".Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
"device evaluation: analysis of the returned device identified: weight to spec, opening curved on anterior, crease fold and black particles in the shell.A visual and microscopic analysis was performed which identified: striated opening on anterior.Based on the device analysis the final assessment is: a striated opening on anterior assessed as surgical damage.Additional, changed, and/or corrected data: d9, h3, h6.
|
| |
|