Model Number 15
Medical Device Problem Code Unexpected Shutdown (4019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/31/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer contacted stryker to report that their device unexpectedly powered off during a patient event.There were no reports of any adverse effects to the patient as a result of the reported issue.
 
Additional Manufacturer Narrative
Stryker contacted the customer to request additional information on the patient.No response has been received from the customer.Patient fields in which information was not provided were intentionally left blank.Stryker performed an initial evaluation of the customer's device and downloaded the device's electronic records from the event for review and was able to verify the reported issue.Stryker continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
 
Additional Manufacturer Narrative
Stryker replaced the device's system pcb assembly as a precaution.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.The cause of the reported issue could not be determined.
 
Event or Problem Description
The customer contacted stryker to report that their device unexpectedly powered off during a patient event.There were no reports of any adverse effects to the patient as a result of the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK® 15 DEFIBRILLATOR/MONITOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
todd bandy
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key13636964
Report Number0003015876-2022-00296
Device Sequence Number11319242
Product Code MKJ
UDI-Device Identifier00883873871669
UDI-Public00883873871669
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K142430
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 10/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number15
Device Catalogue Number99577-001960
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/03/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/03/2022
Supplement Date Received by Manufacturer10/12/2022
Initial Report FDA Received Date03/01/2022
Supplement Report FDA Received Date10/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/21/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-