| Model Number |
15 |
| Medical Device Problem Code |
Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/31/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device unexpectedly powered off during a patient event.There were no reports of any adverse effects to the patient as a result of the reported issue.
|
| |
|
Additional Manufacturer Narrative
|
|
Stryker contacted the customer to request additional information on the patient.No response has been received from the customer.Patient fields in which information was not provided were intentionally left blank.Stryker performed an initial evaluation of the customer's device and downloaded the device's electronic records from the event for review and was able to verify the reported issue.Stryker continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Additional Manufacturer Narrative
|
|
Stryker replaced the device's system pcb assembly as a precaution.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.The cause of the reported issue could not be determined.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device unexpectedly powered off during a patient event.There were no reports of any adverse effects to the patient as a result of the reported issue.
|
| |
|