Model Number 501-014
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Perforation of Vessels (2135)
Date of Event 01/31/2022
Type of Reportable Event Serious Injury
Event or Problem Description
A lead extraction procedure commenced to remove a right ventricular (rv) lead to upgrade a dual chamber pacemaker to a crtd system.The patient was a congenital heart patient and did not have normal anatomy.A venogram showed venous occlusion and an unidentifiable (congenital) large vertical venous structure from the access point (subclavian) at the pocket, moving inferior.There was also a remnant of a previously extracted lead (date and model unknown) in the subclavian vein.Spectranetics lead locking devices (llds) were inserted into the ra and rv leads to provide traction.First, the physician attempted to remove the rv lead using a spectranetics 14f glidelight laser sheath and a spectranetics medium visisheath dilator sheath.With these devices, advancement was made to the superior vena cava (svc) region, but the rv lead was found to be adhered to the ra lead, and both leads pulled back and bunched up together in the right atrium and svc regions.The glidelight device could no longer be advanced.A wire was advanced through the glidelight device to obtain access for re-implantation of new leads.The ra lead was snared from the groin and attempt was made again to remove the ra removed with the 14f glidelight and medium visisheath devices, but the 14f glidelight could not pass the lead on lead binding site.There was much difficulty every time the 14f glidelight device passed the remnant lead present in the subclavian, as there was a lot of resistance.The physician then attempted to remove the rv lead, upsizing to a 16f glidelight and a large visisheath device.Tension was placed on the ra lead from the groin and significant resistance was encountered with the large visisheath in the subclavian area.The physician removed the devices from the patient, with the plan to proceed without the visisheath, but the patient started bleeding from the pocket site (subclavian vein) (mdr #).The bleeding could not be controlled, and the patient's blood pressure dropped.A figure of 8 intervention was attempted by the physician but was unsuccessful due to the amount of blood.The surgeon was paged and was in the room within approximately 5 minutes.Transesophageal echocardiography (tee) showed no effusion and an x-ray showed no blood around the lungs.Rescue interventions were implemented.The surgeons had a difficult time accessing the bleeding site (subclavian vein) as it was under the patient¿s clavicle but was eventually able to control the bleeding by suturing above the bleeding site.They were unsure if this was a sufficient repair, so the surgeons stayed scrubbed in while the physician re-implanted a new ra and rv lead.The surgeons advised to implant as quickly as possible so an lv lead was not attempted.The physician chose to abandon both the rv lead (mdr #1721279-2022-00035) and the ra lead (mdr #1721279-2022-00036), cutting and capping both leads/llds which remained in the patient.The physician did not attempt to unlock either lld prior to cutting and capping.The patient was stable by the end of the procedure.This report captures the large visisheath, the outermost device in the region when the subclavian bleeding (perforation) occurred, requiring intervention.There was no alleged malfunction of any spectranetics devices in use during the procedure.
 
Additional Manufacturer Narrative
Relevant tests/laboratory data unk.The device was discarded, thus no investigation could be completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRANETICS LARGE VISISHEATH DILATOR SHEATH
Common Device Name
DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION
Manufacturer (Section D)
THE SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
Manufacturer (Section G)
THE SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
Manufacturer Contact
barbara creel
9965 federal drive
colorado springs, CO 80921
MDR Report Key13637432
Report Number1721279-2022-00034
Device Sequence Number2174653
Product Code DRE
UDI-Device Identifier00813132020361
UDI-Public(01)00813132020361(17)231215(10)FVM21M15A
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K092378
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/15/2023
Device Model Number501-014
Device Catalogue Number501-014
Device Lot NumberFVM21M15A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/31/2022
Initial Report FDA Received Date03/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
FEMORAL SNARE MANUFACTURER UNK.; SPECTRANETICS 14F GLIDELIGHT LASER SHEATH.; SPECTRANETICS 16F GLIDELIGHT LASER SHEATH.; SPECTRANETICS CVX-300 EXCIMER LASER SYSTEM.; SPECTRANETICS LEAD LOCKING DEVICES.; SPECTRANETICS MEDIUM VISISHEATH DILATOR SHEATH.; ST. JUDE MEDICAL 1888TC RA PACING LEAD.; ST. JUDE MEDICAL 1888TC RV PACING LEAD.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age37 YR
Patient SexFemale
Patient Weight59 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-