| Model Number |
3708640 |
| Medical Device Problem Codes |
Failure to Deliver Energy (1211); High impedance (1291); Unintended Collision (1429)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/28/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant products: product id: 37603, serial#: (b)(4), implanted: (b)(6) 2021, explanted: (b)(6) 2022, product type: implantable neurostimulator.Other relevant device(s) are: product id: 37603, serial/lot #: (b)(4), ubd: 14-jul-2021, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient had high impedances on open circuits.The patient described that she fell in december and then she was not getting stimulation.Impedances were check and it revealed open circuit on all 4 contacts.The healthcare provider (hcp) decided to replace the battery and extension wire.All impedances are now normal.The issue was resolved with the revision surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
H3.Analysis of the ins (s/n (b)(6) found no significant anomaly.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|