WRIGHTS LANE SYNTHES USA PRODUCTS LLC SPEEDSHIFT COMPRESSION IMPLANT KIT 15X20MM OFFSET 6MM; STAPLE, FIXATION, BONE Back to Search Results
Model Number SE-1520-06
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Complainant part is not expected to be returned for manufacturer review/investigation.Reporter is a j&j employee.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on an unknown date speedshift implant kit was found broken inside the sterile packaging.There was no patient involvement.This report involves one (1) speedshift compression implant kit 15x20mm offset 6mm.This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10: additional narrative: h3, h6: a photo investigation was completed: visual analysis of the provided photo revealed that the device is broken.No other issues were found of the device.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H6: part: se-1520-06; bme lot: bse171018; manufacturing date or release to warehouse date: january 25, 2018.Place of manufacture: biomedical enterprises, san antonio, tx.Lot expiration date: january 02, 2023.The product was returned to us customer quality (cq) for evaluation.The us cq team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the speedshift compression implant kit 15x20mm offset 6mm had a broken prong of the inserter and the implant had been deployed inside the original sealed packaging.No other issues were identified.The dimensional inspection was not performed as this is a known issue addressed under capa-007630.The observed condition of the speedshift compression implant kit 15x20mm offset 6mm was consistent with a design related issue.A functional test was not performed as the device was found to be broken inside the original packaging.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the speedshift compression implant kit 15x20mm offset 6mm was observed to broken and deployed inside the original packaging.The potential impact of the design to the complaint condition is addressed under capa-007630.No further corrective action is deemed necessary at this time.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed the following drawings reflecting the current and manufactured revisions were reviewed: speedshift(tm) implant kit.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPEEDSHIFT COMPRESSION IMPLANT KIT 15X20MM OFFSET 6MM
Common Device Name
STAPLE, FIXATION, BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
BIO MEDICAL ENTERPRISES INC
14785 omicron dr # 205
san antonio TX 78245
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key13637607
Report Number2939274-2022-00728
Device Sequence Number11309341
Product Code JDR
UDI-Device Identifier00810633020425
UDI-Public(01)00810633020425
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K142292
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSE-1520-06
Device Catalogue NumberSE-1520-06
Device Lot NumberBSE171018
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/01/2022
Supplement Date Received by Manufacturer03/24/2022
05/13/2022
Initial Report FDA Received Date03/01/2022
Supplement Report FDA Received Date03/24/2022
06/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/25/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-