| Model Number |
N/A |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/17/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported the cylindrical cuff leak.The event timing was before surgery.There was no harm to delay.No adverse events were reported as a result of this malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint is recorded by zimmer biomet under (b)(4).A follow up/ final report will be submitted once investigation is complete.
|
| |
|
Event or Problem Description
|
|
There was no additional event information.
|
| |
|
Additional Manufacturer Narrative
|
|
Functional testing confirmed there was a leak in the cuff.A bubble determined the leak was along the inseam of the cuff.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Additional Manufacturer Narrative
|
|
This event is recorded by zimmer biomet under (b)(4).
|
| |
|
Event or Problem Description
|
|
No additional information is available regarding the incident.
|
| |
|
Search Alerts/Recalls
|