MAUDE Adverse Event Report: ABBOTT MEDICAL AMPLATZER VASCULAR PLUG II; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL AMPLATZER VASCULAR PLUG II; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number 9-AVP2-018
Medical Device Problem Codes Difficult to Fold, Unfold or Collapse (1254); Improper or Incorrect Procedure or Method (2017); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/04/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported a 18mm amplatzer vascular plug ii was chosen for procedure.During the procedure, an attempt to advance the device through a 7fr sheath was made and the user reported feeling resistance.The sheath was exchanged with a larger 8fr sheath and the device was able to be deployed.The user repositioned and re-sheathed the device 3 to 4 times.During the last attempt to deploy this device, while inside the patient, the device presented with a deformation on one of the discs, its shape unspecified.Still attached to the cable, the device was removed and exchanged with an unknown size avpi device and the procedure was successful.The patient did remain stable throughout the procedure.There was no clinically significant delay in procedure and no adverse patient effects.No additional information has been provided.
 
Additional Manufacturer Narrative
An event of advancement difficulty and device deformation was reported.A returned device assessment could not be performed as the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.There were no complaints associated with any other devices from the lot.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.Please note, per the amplatzer vascular plug ii instructions for use, "directions for use: if device position is unsatisfactory: stabilize the wire and re-advance the delivery catheter until the device is completely within the catheter.Reposition and deploy, or remove the device from the patient." d6a & d6b: added implant & explant date.H6: medical device problem code added: 2920 difficult to advance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER VASCULAR PLUG II
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL REG# 2135147
5050 nathan ln n
plymouth MN 55442
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key13638446
Report Number2135147-2022-00069
Device Sequence Number5075630
Product Code KRD
UDI-Device Identifier00811806010427
UDI-Public00811806010427
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K071699
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2024
Device Model Number9-AVP2-018
Device Catalogue Number9-AVP2-018
Device Lot Number7031403
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/04/2022
Supplement Date Received by Manufacturer04/13/2022
Initial Report FDA Received Date03/01/2022
Supplement Report FDA Received Date05/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/05/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-