| Model Number |
383576 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/03/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A sample is available for evaluation.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported while using nexiva¿ closed iv catheter system ¿ dual port maxzero needleless connector a the needle detached from the safety mechanism.There was no report of patient impact.The following information was provided by the initial reporter: material no.: 383576 batch no.: 0300545 it was reported that needle was exposed and not protected, after iv was started.
|
| |
|
Event or Problem Description
|
|
It was reported while using nexiva¿ closed iv catheter system ¿ dual port maxzero needleless connector a the needle detached from the safety mechanism.There was no report of patient impact.The following information was provided by the initial reporter: material no.: 383576 batch no.: 0300545.It was reported that needle was exposed and not protected, after iv was started.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: investigation summary: since no photos or samples displaying the condition reported were available for examination, we were unable to fully investigate this incident.A device history record review showed no non-conformances associated with this issue during the production of this batch.
|
| |
|
Search Alerts/Recalls
|