• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GE MEDICAL SYSTEMS, LLC SIGNA PIONEER; NUCLEAR MAGNETIC RESONANCE IMAGING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GE MEDICAL SYSTEMS, LLC SIGNA PIONEER; NUCLEAR MAGNETIC RESONANCE IMAGING Back to Search Results
Device Problems Use of Device Problem (1670); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Burn(s) (1757); Erythema (1840)
Event Date 12/01/2021
Event Type  Injury  
Event Description
It was reported that after a lumbar spine exam, a patient felt warming on their right forearm.The area developed a 1-2cm area of redness that was treated with wound dressing.The patient had to see a plastic surgeon to have the site debrided as it was not healing.
 
Manufacturer Narrative
Medical device unique identifier: (b)(4).There are no additional device identification numbers.Ge healthcare's investigation is ongoing.A follow up report will be submitted once the investigation has been completed.Device evaluation anticipated, but not yet begun.
 
Manufacturer Narrative
H3: the investigation by ge healthcare (gehc) has been completed.The mr system was operating within specifications and all safety mitigating devices were functional when checked by the gehc field engineer.The root cause of the injury was determined to be inadequate patient padding for the mri procedure.The operator documentation and the user interface describe the appropriate safety measures for padding patients for mr exams.The mr operator has the final responsibility for the use and placement of non-conductive mr compatible padding and preparation of the patient, prior to starting the mr exam procedure.No further actions are planned by gehc.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIGNA PIONEER
Type of Device
NUCLEAR MAGNETIC RESONANCE IMAGING
Manufacturer (Section D)
GE MEDICAL SYSTEMS, LLC
3200 n grandview blvd.
waukesha, WI 53188
Manufacturer (Section G)
GE MEDICAL SYSTEMS, LLC
3200 n grandview blvd.
waukesha, WI 53188
Manufacturer Contact
andy koch
3200 n. grandview blvd.
waukesha, WI 53188
MDR Report Key13638885
MDR Text Key286431239
Report Number2183553-2022-00004
Device Sequence Number1
Product Code LNH
Combination Product (y/n)N
Reporter Country CodeSF
PMA/PMN Number
K153003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/01/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/13/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/04/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age53 YR
Patient SexMale
Patient Weight185 KG
-
-