• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS, INC. ADVIA CENTAUR BR 27.29 (BR); BR 27.29 (BR) IMMUNOASSAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS, INC. ADVIA CENTAUR BR 27.29 (BR); BR 27.29 (BR) IMMUNOASSAY Back to Search Results
Model Number N/A
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/04/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported advia centaur xp cancer antigen 27.29 (br 27.29 (br) results of >450 u/ml were obtained from two samples from the same patient and questioned by the physician(s).The samples were auto diluted and gave over dilution errors.The two samples were repeated with the same instrument with manual dilutions, resulting lower.It is unknown which results from the, advia centaur xp instrument were reported as the corrected result to the physician(s).There are no reports that treatment was altered or prescribed or adverse health consequences due to the discordant, advia centaur xp br 27.29 results.
 
Additional Manufacturer Narrative
A united states customer reported advia centaur xp cancer antigen 27.29 (br 27.29 (br) results of >450 u/ml were obtained from two samples from the same patient and questioned by the physician(s).The samples were auto diluted and gave over dilution errors.The two samples were repeated with the same instrument with manual dilutions, resulting lower.The customer has 2 samples from a patient with multiple myeloma and malignant neoplasma that recovered >450 u/ml with advia centaur br (ca 27.29) kit lot 260 when not diluted but upon 1/5 manual dilution recovered < 9 u/ml.Auto dilution of the samples gave over dilution errors.Siemens reviewed the calibration and control data provided and there is no evidence of a problem.The customer was not able to provide a list of medications/supplements the patient was taking.There is no sample that can be sent to siemens healthineers for evaluation.The patient sample had high levels of immunoglobulin g (igg) which is expected for a person with multiple myeloma.Some of the igg antibodies may be interfering with the advia centaur br (ca 27.29) assay.As stated in the limitation section of the advia centaur xp/xpt br instruction for use (ifu) (11206409) "heterophilic antibodies in human serum can react with reagent immunoglobulins, interfering with in vitro immunoassays.".Dilutions are a way to help identify interference which explains why the diluted results did not match the undiluted results.The cause of the discrepant results seen by the customer with samples from this one patient when using advia centaur br kit lot 260 could not be determined but siemens cannot rule out pre-analytical factors or a sample issue.Based on the investigation, no product problem was identified.The customer is operational.No further action is needed.Mdr 1219913-2022-00074 and mdr 1219913-2022-00075 were filed for different test dates.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVIA CENTAUR BR 27.29 (BR)
Common Device Name
BR 27.29 (BR) IMMUNOASSAY
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS, INC.
511 benedict avenue
tarrytown NY 10591 5097
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS, INC
333 coney street
east walpole MA 02032
Manufacturer Contact
barry memishian
333 coney street
east walpole, MA 02032
5082985306
MDR Report Key13638950
Report Number1219913-2022-00074
Device Sequence Number16688028
Product Code MOI
UDI-Device Identifier00630414196664
UDI-Public00630414196664
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K982680
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/19/2022
Device Model NumberN/A
Device Catalogue Number10334837
Device Lot Number260
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/04/2022
Initial Report FDA Received Date03/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/20/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-