MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPICC SOLO CATHETER WITH SHERLOCK 3CG TPS STYLET 5F; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPICC SOLO CATHETER WITH SHERLOCK 3CG TPS STYLET 5F; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Medical Device Problem Code Material Frayed (1262)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of reft3500 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported on friday, (b)(6) 2022, about 5pm, during the insertion of a picc catheter, there was an occurrence during its use.Picc catheter was attempted to be placed in adult patient (above 60 years old), impatient in the facility.After several unsuccessful attempts to advance the catheter until its mark zero, the catheter was removed from the vein to restart the procedure from the beginning, and then it was noticed that its distal end (the last 5-7 cm) was with memory (an "addicted" position).So, the hydrophilic stylet was removed as well until its end so it could be evaluated, and it was presenting a fracture in its distal end.Initially, the catheter was cut in 43cm.However, in the firsts insertion attempts it was not advancing till its end, and there was always 11cm remaining outside.Due to the advancing failures, the entire catheter was removed and evaluated throughout its length, and at this moment it was identified the memory, at the same moment the stylet was removed to be evaluated.The second cut of the catheter (more 11cm were cut away), was performed after the removal of the stylet.It was evaluated more closely and it was confirmed the fracture of the hydrophilic stylet, so it was removed and no longer used to catheter placement.Customer highlights that prior the catheter insertion attempt, it was performed a visual inspection and fulfill of the catheter's prime with saline solution 0.9%.The catheter was kept aligned on the tray and no issue/memory was observed in it.Additionally, for the catheter cut, it was performed the removal of the stylet keeping outside 2/3 of its length, assuring the stylet was not at the point the catheter was cut.It was observed that the hydrophilic stylet was untouched after the catheter was cut, attending to the needing of keeping it so its technology could be entirely used (by surfing and confirmation of intracavitary catheter tip).The catheter insertion was performed without the hydrophilic stylet, so the patient, users, and professionals would not be under any risk.
 
Event or Problem Description
It was reported on (b)(6) 2022, about 5pm, during the insertion of a picc catheter, there was an occurrence during its use.Picc catheter was attempted to be placed in adult patient (above 60 years old), impatient in the facility.After several unsuccessful attempts to advance the catheter until its mark zero, the catheter was removed from the vein to restart the procedure from the beginning, and then it was noticed that its distal end (the last 5-7 cm) was with memory (an "addicted" position).So, the hydrophilic stylet was removed as well until its end so it could be evaluated, and it was presenting a fracture in its distal end.Initially, the catheter was cut in 43cm.However, in the firsts insertion attempts it was not advancing till its end, and there was always 11cm remaining outside.Due to the advancing failures, the entire catheter was removed and evaluated throughout its length, and at this moment it was identified the memory, at the same moment the stylet was removed to be evaluated.The second cut of the catheter (more 11cm were cut away), was performed after the removal of the stylet.It was evaluated more closely and it was confirmed the fracture of the hydrophilic stylet, so it was removed and no longer used to catheter placement.Customer highlights that prior the catheter insertion attempt, it was performed a visual inspection and fulfill of the catheter's prime with saline solution 0.9%.The catheter was kept aligned on the tray and no issue/memory was observed in it.Additionally, for the catheter cut, it was performed the removal of the stylet keeping outside 2/3 of its length, assuring the stylet was not at the point the catheter was cut.It was observed that the hydrophilic stylet was untouched after the catheter was cut, attending to the needing of keeping it so its technology could be entirely used (by surfing and confirmation of intracavitary catheter tip).The catheter insertion was performed without the hydrophilic stylet, so the patient, users, and professionals would not be under any risk.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a damaged stylet was confirmed; however, the root cause was not identified.The product returned for evaluation was four photographs which depicted an ecg stylet.All four photographs depicted the distal end of the stylet.Curved shape memory and kinks were evident throughout the plastic-coated region.The tip of the stylet appeared to be broken.The fourth photograph also depicted the distal end of the stylet and the distal broken fragment from within the plastic-coated region.Stylet damage was evident in the submitted photographs; however, inspection of the photographs was insufficient to identify the cause of the damage.Consequently this complaint is confirmed as ¿cause uknown¿ at this time.Potential contributing factors include sharp instrument damage and inserting the tip of the stylet past the catheter tip during catheter insertion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC SOLO CATHETER WITH SHERLOCK 3CG TPS STYLET 5F
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
kelsey erickson
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key13639205
Report Number3006260740-2022-00548
Device Sequence Number14721420
Product Code LJS
UDI-Device Identifier00801741034633
UDI-Public(01)00801741034633
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K091324
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2022
Device Model NumberN/A
Device Catalogue Number1295118
Device Lot NumberREFT3500
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/17/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/14/2022
Supplement Date Received by Manufacturer04/08/2022
Initial Report FDA Received Date03/01/2022
Supplement Report FDA Received Date04/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexFemale
-
-