Model Number M00561610
Medical Device Problem Codes Use of Device Problem (1670); Suction Problem (2170)
Health Effect - Clinical Code Perforation (2001)
Date of Event 01/28/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The complainant was unable to provide the lot number.Therefore, the manufacture and expiration dates are unknown.(b)(4).The complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a captivator endoscopic mucosal resection device was used during a endoscopic mucosal resection (emr) procedure performed on (b)(6), 2022.When the device was opened the blue cap was dropped/lost so the blue cap was missing.The technician used their finger to complete suction loop instead.The device was successfully used in the antrum.When used in the fundus, the band captured muscularis due to too much suction and the band was deployed over the whole wall and was not realized until after it was cut, and a perforation occurred.The procedure was successfully completed with a non-bsc clip device.The patient outcome was reported to be fully recovered.There was no additional intervention required and the patient was not admitted to the hospital beyond standard of care.The patient condition at the conclusion of the procedure was reported to be fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTIVATOR EMR
Common Device Name
ENDOSCOPE AND/OR ACCESSORIES
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key13639856
Report Number3005099803-2022-00986
Device Sequence Number5201164
Product Code FDI
UDI-Device Identifier08714729842699
UDI-Public08714729842699
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K140726
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00561610
Device Catalogue Number50661
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/31/2022
Initial Report FDA Received Date03/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-