Model Number LTV 1000
Medical Device Problem Code Unexpected Shutdown (4019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported to vyaire medical that the lap top 1000 ventilator shutdown while on a transport, but indicated there was no patient involvement associated with the reported event.
 
Additional Manufacturer Narrative
Vyaire medical file identification: (b)(4).The suspect device was returned and evaluation is anticipated but not yet begun.Once a final investigation is complete, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
Results of investigation: the suspect device was returned for investigation.Vyaire medical was able to verify the customer complaint regarding unit shutdown during a transport.The customer confirmed that there was no patient involvement.The reported failure was found to be caused by a drained internal battery.Visually inspected uut (unit under test) found a physically damaged dove tail on the rear weldment along with a physically damaged warning label.During investigation, uut was connected onto a known good ac adapter, the "external power" led illuminated solid green with "charge status" illuminated solid red indicated a drained battery.After several minutes, the "charge status" led illuminating flash ember and solid ember indicated the battery being charge.Uut was then powered on and passed post test with no issues.It underwent vent check and passed alarm test, display test, button test, and leak test.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LTV 1000 VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
VYAIRE MEDICAL
510 technology dr
irvine IL 92618
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
erika bonilla
510 technology dr
irvine, IL 92618
7149227837
MDR Report Key13640875
Report Number2021710-2022-15643
Device Sequence Number2213358
Product Code CBK
UDI-Device Identifier00845873000203
UDI-Public(01)00845873000203(11)19991228
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K051767
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLTV 1000
Device Catalogue Number10130-99
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/07/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/01/2022
Supplement Date Received by Manufacturer05/09/2022
Initial Report FDA Received Date03/01/2022
Supplement Report FDA Received Date06/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/28/1999
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-