• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION GLADIATOR ELITE; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION GLADIATOR ELITE; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 24686
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/24/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The target lesion was located in an arteriovenous fistula.A 7.0 x80, 75cm gladiator elite balloon catheter was advanced for dilatation.However, during procedure, the balloon ruptured.The device was removed and the procedure was completed with another of the same device.There were no complications reported and the patient status was stable.
 
Event or Problem Description
It was reported that balloon rupture occurred.The target lesion was located in the arteriovenous fistula.A 7.0 x80, 75cm gladiator balloon catheter was advanced for dilatation, however, during procedure the balloon ruptured.The procedure was completed with another of the same device.There were no complications reported and the patient is stable.It was further reported that the balloon was inflated once at less working pressure.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: a gladiator device was returned for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.A longitudinal tear was identified in the balloon material.The tear measured 73 mm in length and extended from a position 11 mm proximal of the proximal markerband to 13 mm distal of the distal markerband.An examination of the balloon material and markerbands identified no issues which could potentially have contributed to this complaint.A visual examination observed damage to the tip of the device.A visual and tactile examination found multiple shaft kinks and inner shaft kink.Both markerbands were undamaged and present on the shaft of the device.This concludes the product analysis.
 
Event or Problem Description
It was reported that balloon rupture occurred.The target lesion was located in the arteriovenous fistula.A 7.0 x80, 75cm gladiator balloon catheter was advanced for dilatation, however, during procedure the balloon ruptured.The procedure was completed with another of the same device.There were no complications reported and the patient is stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLADIATOR ELITE
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
pmt 741 persiaran cassia selat
bandarcassia, pulau pinan 14110
MY   14110
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13642432
Report Number2134265-2022-02308
Device Sequence Number11914208
Product Code LIT
UDI-Device Identifier08714729809838
UDI-Public08714729809838
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K132810
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/26/2024
Device Model Number24686
Device Catalogue Number24686
Device Lot Number0027390689
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/24/2022
Supplement Date Received by Manufacturer03/01/2022
03/31/2022
Initial Report FDA Received Date03/01/2022
Supplement Report FDA Received Date03/09/2022
04/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/27/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age49 YR
Patient SexMale
Patient Weight73 KG
-
-