Model Number 82-3832
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 01/20/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event or Problem Description
A facility reported a patient was sent to hospital on (b)(6), 2021 after a car accident.On (b)(6), 2021, a hakim valve (id 823832) and a bactiseal catheter (id 823072) were implanted due to hydrocephalus.During a ct check on (b)(6) 2022, the physician found that the hakim valve was occluded.Then they conducted a revision procedure and changed with a new hakim valve (id 823832).The bactiseal catheter remained implanted.
 
Additional Manufacturer Narrative
The hakim valve (id: 823832) was returned for evaluation.Device history record (dhr) - product code 82-3832 with lot: 4610225, showed 1 nr-report: (b)(4) when released to stock.Failure analysis - the valve was visually inspected; no defect was noted.The position of the cam when valve was received was 30 mmh2o.The valve was hydrated.The valve was tested for programming and failed the test.The valve was flushed and failed the test an occlusion was noted.The valve could not be leak tested due to the occlusion.The valve could not be reflux tested due to the occlusion.The cam mechanism was gently moved.The valve was retested for programming and failed the test.The valve could not be pressure tested due to the occlusion.The valve was disassembled.The valve was visually inspected under microscope with the appropriate magnification: biological debris was noted on the ruby ball, on the cam mechanism and on the base plate.The valve passed the test for siphon guard.The root cause for the occlusion problem reported by the customer was due to biological debris was sticking the ruby ball to the seat of the ruby ball.The root cause for the programming issue noted during the investigation is due to biological debris found on the cam mechanism.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROG VALVE INLINE W SG
Common Device Name
CHPV WITH SG
Manufacturer (Section D)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
rue girardet 29
le locle
SZ 
Manufacturer (Section G)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
ch-2400
le locle
SZ  
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key13652808
Report Number3013886523-2022-00071
Device Sequence Number5801481
Product Code JXG
UDI-Device Identifier10886704041825
UDI-Public(01)10886704041825(17)250331(10)4610225
Combination Product (Y/N)N
PMA/510(K) Number
K053107
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2025
Device Model Number82-3832
Device Catalogue Number823832
Device Lot Number4610225
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/15/2022
Supplement Date Received by Manufacturer04/22/2022
Initial Report FDA Received Date03/02/2022
Supplement Report FDA Received Date04/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/09/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-