• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. COMP RVS TRAY CO 44MM; SHOULDER, PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. COMP RVS TRAY CO 44MM; SHOULDER, PROSTHESIS Back to Search Results
Model Number 115370
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 02/15/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Concomitant products: arcom xl 44-36 std hmrl brng cat: xl-115363 lot: 874890; comp rvrs shldr glnsp std 36mm cat: 115310 lot: 502220; comp aug mini bsplt w tpr md cat: 110032420 lot: 64298726.Report source: foreign: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up report will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2022-00446, 0001825034-2022-00447, 0001825034-2022-00448.Product not returned.
 
Event or Problem Description
It was reported that the patient was revised approximately 1 year and 2 days post implantation due to a foreign object identified on post op radiographs.The object was removed and all components were removed and replaced.No additional patient consequences were reported.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, d4, g3, g7, h1, h2, h4, h10 if any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Not reportable.- the returned fragment does not match any of the returned implants.This event will be reported under medwatch: 0001825034-2022-01296.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMP RVS TRAY CO 44MM
Common Device Name
SHOULDER, PROSTHESIS
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13653259
Report Number0001825034-2022-00445
Device Sequence Number16736686
Product Code KWS
UDI-Device Identifier00880304543201
UDI-Public(01)00880304543201(17)310219(10)104620
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K193373
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number115370
Device Catalogue Number115370
Device Lot Number104620
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/17/2022
Supplement Date Received by Manufacturer03/02/2022
05/04/2022
Initial Report FDA Received Date03/02/2022
Supplement Report FDA Received Date03/29/2022
05/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/19/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age84 YR
Patient SexFemale
-
-