• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION EFFLUENT BAG; SYSTEM, HEMODIALYSIS, REMOTE ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION EFFLUENT BAG; SYSTEM, HEMODIALYSIS, REMOTE ACCESSORIES Back to Search Results
Catalog Number 107650
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/03/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that an external fluid leak was observed from a 9l effluent bag at the level of the weld, during use.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Additional Manufacturer Narrative
Phone number: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
The actual device was discarded; however, three companion samples were received for evaluation.A visual inspection and leak test were performed, and no issues were observed.The reported condition was not verified.A nonconformance has been opened to address this issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
The reported condition was not verified from the evaluation of the received companion samples.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EFFLUENT BAG
Common Device Name
SYSTEM, HEMODIALYSIS, REMOTE ACCESSORIES
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MALTA
a47 industrial estate
marsa
MT  
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key13653335
Report Number1416980-2022-00835
Device Sequence Number5780492
Product Code MON
UDI-Device Identifier07332414075828
UDI-Public(01)07332414075828
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2022
Device Catalogue Number107650
Device Lot Number21H23V490
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/09/2022
Supplement Date Received by Manufacturer03/07/2022
07/06/2022
Initial Report FDA Received Date03/02/2022
Supplement Report FDA Received Date04/04/2022
07/28/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-