| Model Number |
X SERIES |
| Medical Device Problem Code |
Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Complainant alleged that during functional testing, the device failed self-test for defib and pacer functions.Complainant did not indicate that there was any patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer was contacted for return of the suspect product.The customer has responded and indicated the product will not be returning to zoll.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation; the customer's report was observed and attributed to a system interconnection flex cable that was not properly seated to a connector.The cable was reseated to correct the reported problem.The device was recertified and returned to the customer.It could not be firmly established what caused the mis-alignment.Analysis for reports of this type has not identified an increase in trend.
|
| |
|