C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRICATH® FOLEY CATHETER Back to Search Results
Model Number 0167L22
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/07/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the patient heard a loud pop sound from the foley catheter balloon.Upon examination, nurse immediately checked the balloon of foley catheter and found that it was popped.
 
Event or Problem Description
It was reported that the patient heard a loud pop sound from the foley catheter balloon.Upon examination, nurse immediately checked the balloon of foley catheter and found that it was popped.
 
Additional Manufacturer Narrative
The reported issue was inconclusive as no sample was returned for evaluation.A potential root cause for this failure could be "over inflation during use/ pinhole in balloon or cuff/ wall thickness too thin".It is unknown whether the device had met relevant specifications.The product was used for treatment purposes.It was unknown whether the product had caused the reported failure.A review of the device history record did not show any problem or conditions that would have contributed to the reported issue.Based on the results of the investigation, no additional actions are needed.The device was not returned for evaluation.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: visually inspect the product for any imperfections or surface deterioration prior to use.¿ use luer tip syringe to inflate with stated ml of sterile water.Or ¿ for pre-filled products, remove clip and squeeze reservoir to inflate with stated ml of sterile water.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.For urological use only.Manufacturer: to deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If necessary, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Valve type: use luer slip tip syringe.Do not use needle.Recommended inflation capacities 3cc balloon: use 5ml sterile water 5cc balloon: use 10ml sterile water 15cc balloon: use 20ml sterile water 20cc balloon: use 25ml sterile water 30cc balloon: use 35ml sterile water 40cc balloon: use 45ml sterile water 75cc balloon: use 80ml sterile water do not exceed recommended capacities.The device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRICATH® FOLEY CATHETER
Common Device Name
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key13653540
Report Number1018233-2022-00984
Device Sequence Number2442248
Product Code KOD
UDI-Device Identifier00801741018978
UDI-Public(01)00801741018978
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K922431
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2025
Device Model Number0167L22
Device Catalogue Number0167L22
Device Lot NumberNGEZ2509
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/07/2022
Supplement Date Received by Manufacturer04/18/2022
Initial Report FDA Received Date03/02/2022
Supplement Report FDA Received Date05/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age77 YR
Patient SexMale
-
-