Model Number 1011493-15
Medical Device Problem Codes Difficult to Remove (1528); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/10/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the procedure was to treat an 80% stenosed lesion located in a moderately calcified and difficult angled right renal artery.A hi-torque (ht) spartacore guide wire was placed in position.The herculink stent system had resistance with the guide wire during advancement but the physician got it into position.However, the physician lost access to the artery and the guide wire popped from the artery but did not damage the artery.The stent couldn't be removed from the guide wire so the stent and the guide wire were pulled as one unit.The procedure was aborted due to the anatomical challenges.There was no adverse patient effect and there was no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device was received.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.The spartacore guide wire referenced is filed under a separate medwatch report number.
 
Additional Manufacturer Narrative
A visual inspection was performed on the returned device.The reported difficult to advance/position and difficult to remove were confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.The investigation determined the reported difficulties appear to be related to the operational context of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RX HERCULINK ELITE STENT SYSTEM
Common Device Name
STENT, RENAL
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key13653599
Report Number2024168-2022-02137
Device Sequence Number19462772
Product Code NIN
UDI-Device Identifier08717648078019
UDI-Public08717648078019
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
P110001
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2022
Device Model Number1011493-15
Device Catalogue Number1011493-15
Device Lot Number0010661
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/15/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/10/2022
Supplement Date Received by Manufacturer03/07/2022
Initial Report FDA Received Date03/02/2022
Supplement Report FDA Received Date03/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/06/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
HI-TORQUE SPARTACORE GUIDE WIRE.
Patient Age82 YR
Patient SexFemale
Patient Weight47 KG
-
-