• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH MAGELLAN 3MLSFTY COMB 22X1 1/2; NEEDLE, HYPODERMIC, SINGLE LUMEN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH MAGELLAN 3MLSFTY COMB 22X1 1/2; NEEDLE, HYPODERMIC, SINGLE LUMEN Back to Search Results
Model Number 8881833215
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/11/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported that they were giving the patient winrho im and as the writer pushed the plunger on the syringe, medication shot out where the syringe met the needle.The syringe was checked prior to giving the dose and was connected correctly.They were unsure how much medication the patient received, so they had to give a second dose.The leak occurred at the gray area.No injuries occurred.
 
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Additional Manufacturer Narrative
The device history record (dhr) was reviewed and indicated that the product was released accomplishing all quality standards.Prior to a lot¿s release, the lot must be deemed acceptable by-passing inspections that are based on a valid sampling plan.Inspectors routinely examine a statistical sample both physically and visually.Specifically, samples are inspected for leaking.The shop orders used for this lot met all defined acceptance requirements and were released.One decontaminated sample was received at the manufacturing site for the investigation.Visual and functional evaluations were performed, and the reported condition could not be confirmed; the sample was tested for leaking and no issues were observed.A root cause could not be determined based on all available information as the reported issue was not confirmed.There were no anomalies found during a review of the dhr¿s, maintenance records, or process monitoring data.There was no evidence of any systemic failure of the manufacturing process.Based on the investigation and root cause analysis, a corrective action is not applicable at this time.Functional testing and visual inspections are being performed according to our current quality standards and inspection procedures.This complaint will be utilized for tracking and trending purposes to determine the need for corrective actions.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGELLAN 3MLSFTY COMB 22X1 1/2
Common Device Name
NEEDLE, HYPODERMIC, SINGLE LUMEN
Manufacturer (Section D)
CARDINAL HEALTH
2010 east international speedw
deland FL 32724
Manufacturer (Section G)
CARDINAL HEALTH
2010 east international speedw
deland FL 32724
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key13653765
Report Number1017768-2022-00999
Device Sequence Number2481645
Product Code FMI
UDI-Device Identifier10192253024722
UDI-Public10192253024722
Combination Product (Y/N)N
Initial Reporter CountryCA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8881833215
Device Catalogue Number8881833215
Device Lot Number010467
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/22/2022
Supplement Date Received by Manufacturer02/22/2022
Initial Report FDA Received Date03/02/2022
Supplement Report FDA Received Date06/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/09/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-