MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES NAVIO HANDPIECE; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES NAVIO HANDPIECE; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number PFSR110137
Medical Device Problem Codes Break (1069); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: case: (b)(4).
 
Event or Problem Description
It was reported that when they went to bur the femoral condyle during a navio assisted surgery, the first anspach drill wouldn't spin (the handpiece moved the drill forward but the drill wouldn't spin).They checked the surgeon was pressing the black pedal and were on "cut femur", unplugged the drill and tried again but the drill wouldn't work.When a second drill and long attachment, and the same burr were added into the first handpiece, they had not been put together properly, so they failed calibration.Upon removing the second drill from the first handpiece, the second long attachment got stuck in the first handpiece and couldn't be removed.Therefore, the surgeon was repeatedly hit with a mallet, breaking the first handpiece, and potentially damaging the second long attachment and second drill.They opened a second handpiece and continued using the previous drill 2 and long attachment 2, but the nurse found it difficult to insert the burr into long attachment (previously damaged with the surgeon's hammering), so the surgeon hammered it in.The handpiece then passed calibration and the surgeon started to use it to remove bone, but at some point, the burr detached from the second drill and wouldn't stay inside the drill even when assembled correctly (previously damaged with the surgeon's hammering).Finally, the surgery was completed satisfactorily using a third handpiece, drill, and long attachment with a significant delay.The patient was not harmed beyond the problem reported.After surgery, during testing, it was found that the burr wouldn't go into the first long attachment.
 
Additional Manufacturer Narrative
Results of investigation: the navio handpiece, pfsr110137, s/n(b)(6), used for treatment was returned for evaluation.A relationship between the reported event (physical resistance/sticking) and the device could not be established due to damaged/broken parts.Nothing was identified visually that contributed to the reported failure.A functional evaluation could not be completed due to damaged/broken parts.The snap lock is in pieces.The second issue (break) was visually confirmed.Although the reported problem was not confirmed, the most likely cause of this event is improper handling of the handpiece.In the event the drill/long attachment is stuck in the handpiece, do not use excessive force to remove the drill/long attachment.It is most likely stuck due to lack of lubrication that is to be applied to the snap lock prior to sterilization.Place the drill/long attachment/handpiece into the sterilizer.Following cooling, the parts will separate.Lubricate the snap lock prior to sterilization to minimize this event.The navio surgical system user's manual provides instructions for proper set up of the device and accessories.Care and caution should be exercised during the surgical site setup, surgery and tear down to protect the device from an unforeseen force, or damage.Refer servicing only to trained smith & nephew service personnel.Equipment damage or personal injury may result if the system is not serviced properly.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.A historical escalation event review was completed.The review determined that prior escalation actions are applicable to the scope of this complaint however no further escalation action is required.The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still adequate.Although no further containment or corrective action is recommended or required at this time, the failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIO HANDPIECE
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13653812
Report Number3010266064-2022-00157
Device Sequence Number2369472
Product Code OLO
UDI-Device Identifier00885556628515
UDI-Public00885556628515
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K191223
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPFSR110137
Device Catalogue NumberPFSR110137
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/11/2022
Supplement Date Received by Manufacturer04/28/2022
Initial Report FDA Received Date03/02/2022
Supplement Report FDA Received Date05/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/12/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ANSPACH EMAX 2 PLUS HAND PIECE, SN(B)(6); ANSPACH EMAX 2 PLUS HAND PIECE, SN(B)(6); LONG ATTACHMENT,SN(B)(6); LONG ATTACHMENT,SN(B)(6); NAVIO SURGICAL SYSTEM UK,SN(B)(6)
Patient SexUnknown
-
-