• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON MEDICAL SYSTEMS BD SAFETYGLIDE¿ SHIELDING HYPODERMIC NEEDLE; HYPODERMIC SINGLE LUMEN NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON MEDICAL SYSTEMS BD SAFETYGLIDE¿ SHIELDING HYPODERMIC NEEDLE; HYPODERMIC SINGLE LUMEN NEEDLE Back to Search Results
Model Number 305924
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/25/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
There were multiple medical device types and 510k¿s reported to be involved.The information for the additional info is as follows: medical device type: fmf common device name: piston syringe.Pma / 510(k)#: k980987.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the plunger rod was broken on the bd safetyglide¿ shielding hypodermic needle.There was no report of patient impact.The following information was provided by the initial reporter: ¿syringe plunger does not have plunger top for finger to push vaccine in with plunger.
 
Event or Problem Description
It was reported that the plunger rod was broken on the bd safetyglide¿ shielding hypodermic needle.There was no report of patient impact.The following information was provided by the initial reporter: syringe plunger does not have plunger top for finger to push vaccine in with plunger.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval yes.D10: returned to manufacturer on: 24-mar-2022.Investigation summary: one photo and one sample of a 3ml luer-lock syringe with open package from batch 9283342 were received.Through visual inspection, it was observed that thumbrest is missing as it is completely broken off from plunger.The condition observed is non-conforming per product specification.Potential root cause for the broken thumbrest defect is associated with the assembly process.These conditions are occurring at/below their expected frequency.Therefore, no corrective action is required at this time.Batch 9283342 is considered in compliance with our product specification requirements.A device history record review showed no rejected inspections or quality issues during the production of the provided batch number 9283342 that could have contributed to the reported defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD SAFETYGLIDE¿ SHIELDING HYPODERMIC NEEDLE
Common Device Name
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer (Section G)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13654222
Report Number1213809-2022-00110
Device Sequence Number20284036
Product Code FMI
UDI-Device Identifier30382903059240
UDI-Public30382903059240
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K951254
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date09/30/2024
Device Model Number305924
Device Catalogue Number305924
Device Lot Number9283342
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/17/2022
Supplement Date Received by Manufacturer04/01/2022
Initial Report FDA Received Date03/02/2022
Supplement Report FDA Received Date04/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/10/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-