• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: K2M, INC. CAYMAN UNITED PLATE INSERTER; INTERVERTEBRAL FUSION DEVICE WITH INTEGRATED FIXATION, LUMBAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

K2M, INC. CAYMAN UNITED PLATE INSERTER; INTERVERTEBRAL FUSION DEVICE WITH INTEGRATED FIXATION, LUMBAR Back to Search Results
Model Number 7902-90033
Medical Device Problem Codes Mechanical Jam (2983); Failure to Eject (4010)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/31/2022
Type of Reportable Event Malfunction
Event or Problem Description
A company representative reported that a cayman united inserter would not release from the implant during surgery.The procedure was completed successfully with no surgical delay and no adverse consequence to the patient.
 
Additional Manufacturer Narrative
Device history records were reviewed and no relevant manufacturing issues were identified.Complaint history was reviewed and no similar complaints for the reported lot number were identified.The returned device was visually inspected.The tip of the inserter was not correctly aligned and the space between the two gripping tips of the device was tight.Additionally, the distal end of the inserter shows marking from malleting.The device was functionally tested and the reported event was confirmed.Release of a sample plate was more difficult with the complaint inserter than with the sample inserter.It is determined that the device was not correctly assembled by the user following reprocessing and the incorrect alignment of the tip caused difficulty disengaging the device.The root cause of the event is misassembly of the device by the user during reprocessing.
 
Event or Problem Description
A company representative reported that a cayman united inserter would not release from the implant during surgery.The procedure was completed successfully with no surgical delay and no adverse consequence to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAYMAN UNITED PLATE INSERTER
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH INTEGRATED FIXATION, LUMBAR
Manufacturer (Section D)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer (Section G)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer Contact
rita karan
2 pearl court
allendale, NJ 07401
2017608000
MDR Report Key13654428
Report Number3004774118-2022-00073
Device Sequence Number17587429
Product Code OVD
UDI-Device Identifier10888857383708
UDI-Public10888857383708
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K172941
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7902-90033
Device Catalogue Number7902-90033
Device Lot NumberJYYG
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/31/2022
Supplement Date Received by Manufacturer04/18/2022
Initial Report FDA Received Date03/02/2022
Supplement Report FDA Received Date05/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/19/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexMale
Patient Weight200 KG
-
-