HOLOGIC, INC SELENIA DIMENSIONS MAMMOGRAPHY SYSTEMS, 3D Back to Search Results
Model Number SDM-SYS-9000-3D
Medical Device Problem Code Unintended System Motion (1430)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The c-arm switches were not returned to hologic for evaluation.The rotary switch was returned and functional testing revealed the switch was functioning as intended.Due to the lack of information provided and all the parts not being returned for evaluation, a definitive root cause cannot be determined.Hologic strives for continuous improvement and will continue to monitor and trend for safety.
 
Event or Problem Description
It was reported that during cleaning of the unit after a patient exam, the user pressed the auto correction button and the unit continued motion to -45 degrees.No injury reported.A field engineer was dispatched to the site and determined the c-arm switches and rotary switch needed to be replaced.Once this was completed the system was working as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELENIA DIMENSIONS MAMMOGRAPHY SYSTEMS, 3D
Common Device Name
MAMMOGRAPHY SYSTEM
Manufacturer (Section D)
HOLOGIC, INC
36 & 37 apple ridge road
danbury CT 06810
Manufacturer Contact
kristin fornieri
36 & 37 apple ridge road
danbury, CT 06810
2037318491
MDR Report Key13655369
Report Number1220984-2022-00010
Device Sequence Number5973478
Product Code OTE
UDI-Device Identifier15420045510784
UDI-Public(01)15420045510784(11)180106
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P080003
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSDM-SYS-9000-3D
Device Catalogue NumberSDM-SYS-9000-3D
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/01/2022
Initial Report FDA Received Date03/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/06/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-