Model Number FV441T
Medical Device Problem Codes Mechanical Problem (1384); Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 01/12/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturing and quality control data: the progav® shunt system was manufactured by a qualified employee in may 2021.Deviations during assembly did not occur.The system was sterilized by miethke and released for shipment after final inspection.For the purposes of this investigation, the progav® 2.0 shunt system is comprised of a progav® adjustable pressure valve with a normal pressure range of 0 to 20 cmh2o and a shuntassistant® fixed pressure valve with a fixed pressure range of 20 cmh2o.The shunt system was inspected as article fv441t.All parameters have been inspected and approved during the manufacturing process.All parameters have been assessed as meeting specifications.Visual inspection: no significant deformations or damage of the valves were detected during the visual inspection.Permeability test: a permeability test has shown that both valves are permeable.Adjustment test: the progav® was tested and is adjustable to all specified pressures.Braking/klick force and brake function test: the brake functionality test has shown that the brake function is fully operational and the braking force is within the given tolerances.Internal inspection: after dismantling of the valves, minimal deposits were found in shuntassistant®.Results: based on our investigation, we are unable to substantiate the clinical claim of occlusion and of non-adjustability.At the time of our investigation, the valves were shown to be permeable.However, it is possible that the deposits observed inside the valve could have caused the malfunction in the past.As described in scientific literature, the problem encountered is one of the known, inevitable risks of hc-therapy by shunt implants.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke (b)(4).No further regulatory actions are required from our point of view.
 
Event or Problem Description
It was reported that a progav 2.0 shuntsystem (part # fv441t) was implanted during a procedure performed on (b)(6) 2021.According to the complainant, the shunt system was believed to be blocked and was not able to be adjusted.The patient underwent a revision procedure on (b)(6) 2022.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: (b)(6).Height: 65 centimeters (cm).Weight: (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV SYS PAED.WITH .SA 20 A.PRECHAMBER
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key13655971
Report Number3004721439-2022-00038
Device Sequence Number16764573
Product Code JXG
UDI-Device Identifier04041906132545
UDI-Public4041906132545
Combination Product (Y/N)N
Initial Reporter CountryRS
PMA/510(K) Number
K103003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/04/2026
Device Model NumberFV441T
Device Catalogue NumberFV441T
Device Lot Number20053088
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/16/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/03/2022
Initial Report FDA Received Date03/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/10/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age3 MO
Patient SexUnknown
Patient Weight2 KG
-
-