• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LIMITED MANUAL FILL INFANT/PAEDIATRIC HUMIDIFICATION CHAMBER; BTT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LIMITED MANUAL FILL INFANT/PAEDIATRIC HUMIDIFICATION CHAMBER; BTT Back to Search Results
Model Number MR225X
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/31/2022
Type of Reportable Event Malfunction
Event or Problem Description
A healthcare facility in (b)(6) reported via a fisher and paykel healthcare (f&p) field representative that an mr225x manual fill humidification chamber had cracked at the connection port of the chamber which caused a leak.The chamber was being used for 10 days.There was no patient consequence.
 
Additional Manufacturer Narrative
(b)(4).The complaint mr225x manual fill humidification chamber was returned to fisher & paykel healthcare (f&p) where it was visually inspected.Visual inspection of the chamber revealed it was cracked and broken at the port.We are unable to determine what caused the crack on the chamber dome.It should be noted the complaint mr225 chamber was used for ten days.Every mr225 chamber is pressure tested following the manufacturing process to check for any leaks present in the chamber dome due to cracks and other causes.Any chamber which fails this test is rejected.The user instructions that accompany the mr225 humidification chamber specify in the warning section: do not use beyond 7 days maximum duration of use.Do not soak, wash or sterilize this product.Avoid contact with chemicals, cleaning agents or hand sanitizers.Check all connections are tight.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MANUAL FILL INFANT/PAEDIATRIC HUMIDIFICATION CHAMBER
Common Device Name
BTT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LIMITED
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LIMITED
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
173 technology dr.
suite 100
irvine, CA 92618
9494534000
MDR Report Key13656076
Report Number9611451-2022-00172
Device Sequence Number2455830
Product Code BTT
UDI-Device Identifier09420012401519
UDI-Public(01)09420012401519(10)2101419729(11)201203
Combination Product (Y/N)Y
Initial Reporter CountryAS
PMA/510(K) Number
K934140
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 02/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMR225X
Device Catalogue NumberMR225X
Device Lot Number2101419729
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/16/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/01/2022
Initial Report FDA Received Date03/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/03/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-